Keysight Technologies CX3300 Series User Manual page 35

Current waveform analyzer 2- and 4-channel models
Table of Contents

Advertisement

Figure 1-8
No
Replace
CX1104A.
Current sensor and/or
resistive sensor head is
defective. Replace them or
contact Keysight sales office
for further troubleshooting.
Keysight CX3300 User's Guide, Edition 5
CX1100 Troubleshooting
Yes
Do you have a
"working" resistive
sensor head?
Yes
Replace
the resistive sensor head
with the "working" one.
Perform
Fail
User Calibration.
Pass
The resistive sensor
head is defective.
Replace it.
START
Is it
CX1104A?
No
Is it
CX1103A, CX1105A, or
CX1151A?
No
Is it
CX1152A?
No
Do you have
a "working" CX1203A?
Yes
Replace the sensor head with
the "working" CX1203A.
(set the slide switch to 0 )
Perform
User Calibration.
Pass
The sensor head
is defective.
Replace it.
Setting Up the CX3300
Maintenance
Yes
Replace
CX1103A, CX1105A,
or CX1151A.
Yes
Go to CX1152A
troubleshooting.
No
No
Current sensor and/or
sensor head is defective.
Replace them or
contact Keysight sales office
for further troubleshooting.
Fail
Replace CX1100.
35