Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Figure 3-3
94
4. Open the Display tool palette and drag and drop the A-B marker icon
on the waveform.
5. Put the X1 line on the 200 s position.
6. Put the X2 line on the 800 s position.
7. Read the marker Y1 and Y values and record them as follows.
Ioff_positive = Y
Ioff_offset = Y1
Positive Pulse Acquisition
d. Press the High Reso key to set the high resolution mode ON. The High
Reso key is located in the Horizontal control on the CX3300 front panel.
e. Control the CX3300 and acquire the waveform as same as the step 3-c.
And read the marker Y1 and Y values then record them as follows.
Ion_positive = Y
Ion_offset = Y1
f. Disable the current source (Keysight B2900) output.
X1 line
Y1
Y
Keysight CX3300 User's Guide, Edition 5