Planmeca FIT CAD System User Manual page 186

Table of Contents

Advertisement

22 EMC information
Immunity test
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
178
Planmeca FIT CAD/CAM system
IEC 60601 Test level
3 Vrms 150 kHz to 80
MHz
3 V/m 80.0 MHz to 2.5
GHz
Compliance level
Not applicable
3 V/m
Electromagnetic
environment -guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the Planmeca
PlanScan scanner,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d = 1.2√P
d = 1.2√P 80 MHz to
800 MHz
d = 2.3√P 800 MHz to
2.5. GHz
where P is the
maximum output power
rating of the transmitter
in watts (W) according
to the transmitter
manufacturer and d is
the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
surveya should be less
than the compliance
level in each frequency
rangeb.
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
User's manual