22 EMC information
Immunity test
Voltage dips, short
interruptions and
variations on power
supply input lines IEC
61000-4-11
Power frequency
(50/60Hz)
Magnetic field IEC
61000-4-8
Note:Uт is the A.C. mains voltage prior to application of the test level.
Conducted RF IEC
61000-4-6
174
Planmeca FIT CAD/CAM system
IEC 60601 Test level
<5% Uт (>95% dip in
Uт) for 0.5 cycle
40% Uт (60% dip in
Uт) for 5 cycle
70% Uт (30% dip in
Uт) for 25 cycle
3.0 A/m
3 Vrms 150 kHz to 80
MHz
Compliance level
Not applicable
3.0 A/m
Not applicable
Electromagnetic
environment -guidance
Main power quality
should be that of a
typical commercial or
hospital environment. If
the user of the
BSVD-1000 system
requires continued
operation during power
mains interruptions, it is
recommended that the
Planmeca Emerald
scanner be powered
from an uninterruptible
power supply (UPS) or
a battery.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the Planmeca
Emerald scanner,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
User's manual