Injection Valve; Protection; Inline Sample Preparation; Rinsing The Sample Path - Metrohm 882 Compact IC plus Manual

Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
5.7

Injection valve

5.7.1

Protection

5.8

Inline sample preparation

5.9

Rinsing the sample path

882 Compact IC plus – Anion
To prevent contamination of the injection valve, a 6.2821.120 inline filter
(see Chapter 3.10, page 32) should be mounted between the high pres-
sure pump and the pulsation damper.
To protect the separation column (see Chapter 3.17, page 47) against for-
eign particles which can affect the separating efficiency, we recommend
that all samples undergo a microfiltration (filter 0.45 µm). The ultrafiltra-
tion cell can be used for filtration (see manual of the IC Equipment for
Ultrafiltration).
Matrix-loaded samples (e.g. blood, oil) should be prepared for the meas-
urement by means of dialysis (see manual on the IC Equipment for Dialy-
sis).
If the concentration of the sample is too high, the sample should be dilu-
ted before feeding (see documentation on the IC Equipment for Sample
Dilution).
For an overview of all Metrohm inline sample preparation methods go to
the following
website:http://misp.metrohm.com
Before a new sample can be measured, the sample path must be rinsed
with it so that the measuring result is not falsified by the previous sample
(Sample carry-over).
In the case of automated sample feeding, the rinsing time should be at
least 3 times the transfer time.
The transfer time is the time required by the sample to flow from the sam-
ple vessel to the end of the sample loop. ##NO_MATCH##.
Ascertaining the transfer time
To ascertain the transfer time, proceed as follows:
1 Emptying the sample path
Pump air through the sample path (pump tubing, tubing connec-
tions, sample loop) for several minutes until all liquid is displaced by
the air.
5 Handling and maintenance
■■■■■■■■
69

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents