3.10 Injection valve

1
Connection capillary
Connection to inline filter.
3
Pulsation absorber (6.2620.150)
5
Connection capillary
Connection to injection valve.
3.10
Injection valve
■■■■■■■■
26
Figure 11
Pulsation absorber
The pulsation absorber is completely connected. No installation work is
required.
The injection valve connects the eluent path to the sample path. By a
quick and precise switching of the valve a defined quantity of sample solu-
tion is injected and flushed to the separation column with the eluent.
The quantity of sample solution injected is determined by:
the volume of the sample loop or
by an 800 Dosino when the Metrohm intelligent Partial Loop Injection
Technique (MiPT) is used.
The choice of sample loop depends on the application. The following
sample loops are normally used:
Table 1
Which sample loop do I need?
Application
Cation determination
Anion determination with suppression
Anion determination without suppression
MiPT
2
PEEK pressure screws, short
(6.2744.070)
4
Holder for pulsation absorber
930 Compact IC Flex Oven/ChS (2.930.2200)
■■■■■■■■■■■■■■■■■■■■■■
Sample loop
20 µL
20 µL
100 µL
250 µL