3.10 Injection valve
1
Connection capillary
Connection to inline filter.
3
Pulsation absorber (6.2620.150)
5
Connection capillary
Connection to injection valve.
3.10
Injection valve
■■■■■■■■
26
Figure 11
Pulsation absorber
The pulsation absorber is completely connected. No installation work is
required.
The injection valve connects the eluent path to the sample path. By a
quick and precise switching of the valve a defined quantity of sample solu-
tion is injected and flushed to the separation column with the eluent.
The quantity of sample solution injected is determined by:
the volume of the sample loop or
■
by an 800 Dosino when the Metrohm intelligent Partial Loop Injection
■
Technique (MiPT) is used.
The choice of sample loop depends on the application. The following
sample loops are normally used:
Table 1
Which sample loop do I need?
Application
Cation determination
Anion determination with suppression
Anion determination without suppression
MiPT
2
PEEK pressure screws, short
(6.2744.070)
4
Holder for pulsation absorber
930 Compact IC Flex Oven/ChS (2.930.2200)
■■■■■■■■■■■■■■■■■■■■■■
Sample loop
20 µL
20 µL
100 µL
250 µL