3.9 High pressure pump
1
Pressure screw
For connecting the coupling (14-
pump head inlet capillary (13-
Can be ordered together with the coupling
under the number (6.2744.230).
3
Clamping screw
5
Backup ring
3.9.2

Deaerating the high pressure pump

30
■■■■■■■■
2
) to the
7
).
1 Connecting coupling
Fasten the coupling (14-2) with a pressure screw (14-1) on the pump
head inlet capillary (13-7).
2 Connecting eluent aspiration tubing
Caution
The clamping screws must be tightened carefully. To tighten, grip
the coupling (14-2) with the key (6.2739.000) and grip the clamp-
ing screw (14-3) with the wrench (6.2621.050).
Plug the eluent aspiration tubing (14-4) into the coupling (14-2).
Tighten the clamping screw (14-3).
The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.
Caution
The high pressure pump must not be deaerated before the initial start-
up (see Chapter 4.1, page 50).
Deaerate the high pressure pump as follows (see Figure 15, page 31):
Deaerate the high pressure pump
The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.
2
Coupling (6.2744.230)
For connecting the eluent connection capil-
4
lary (14-
) to the input of the high pressure
pump.
4
Eluent aspiration tubing
Eluent aspiration tubing (6.1834.080) or
(6.1834.090).
■■■■■■■■■■■■■■■■■■■■■■
882 Compact IC plus – Anion

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents