Separation Column; Separating Efficiency - Metrohm 882 Compact IC plus Manual

Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
5.12

Separation column

5.12.1

Separating efficiency

882 Compact IC plus – Anion
If the rotor is correctly inserted, its sealing surface will be approx.
4 mm within the housing. If this is not the case, the rotor must be
moved into the right position from below using a sharp object.
5 Cleaning the new connecting piece
Clean the surface of the new connecting piece (36-2) with etha-
nol using a lint-free cloth.
6 Inserting the new connecting piece
Insert the connecting piece (36-2) into the housing (36-4) in such
a way that the connector 1 is on top and the three pins of the
connecting piece fit into the corresponding recesses on the hous-
ing.
Reattach the union nut (36-1) onto the coupling and tighten by
hand..
7 Connecting and conditioning the suppressor
Reconnect all capillaries of the suppressor to the IC system.
Before switching the suppressor over for the first time, rinse the
three suppressor units with solution for 5 minutes.
Which analysis quality can be attained, depends to a great extent on the
separating efficiency of the separation column used. The separating effi-
ciency of the selected separation column must be sufficient for the analy-
sis problems present. If difficulties occur, you should always first check the
quality of the separation column by recording a standard chromatogram.
You can find detailed information on the separation columns available
from Metrohm in the leaflet provided along with your separation column,
in the Metrohm IC-Column Program (available via your Metrohm
agent) or in the Internet at
Ion chromatography. You can request free information on special IC appli-
cations in the corresponding "Application Bulletins" or "Application
Notes", which are available in the Internet at
in the Applications area or via the Metrohm agent responsible.
http://www.metrohm.com
http://www.metrohm.com
5 Handling and maintenance
in the product area
■■■■■■■■
79

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents