5.11 Metrohm Suppressor Module (MSM)
5.11.3

Maintenance

5.11.3.1
Parts of the suppressor
1
Figure 36
Parts of the suppressor
1
Union nut
3
Rotor
5
Slot in the housing
5.11.3.2
Regenerating the suppressor
74
■■■■■■■■
parable conditions, you should normally work with a freshly regenerated
suppressor unit.
Caution
The suppressor must never be switched over in a dry state, as there is a
risk of jamming. If the suppressor is in a dry state, it must be rinsed for
at least 5 minutes before it may be switched over.
Caution
In the case of reduced capacity or high backpressure, the suppressor
must be regenerated (see Chapter 5.11.3.2, page 74), cleaned (see
Chapter 5.11.3.3, page 76) or replaced (see Chapter 5.11.3.4, page
78).
2
If the suppressor units are loaded for a longer period with certain heavy
metals (e.g. iron) or organic impurities, these can no longer be completely
removed with the standard regeneration solution. The capacity of the sup-
pressor units is consequently affected, which can result in reduced phos-
3
2
Connecting piece (6.2832.010)
4
Housing
■■■■■■■■■■■■■■■■■■■■■■
4
5
882 Compact IC plus – Anion

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents