5.5 High pressure pump

5.4.2

Operation

5.4.2.1
Supply bottle
5.4.2.2
Aspiration filter
5.4.2.3
Changing the eluent
5.5
High pressure pump
5.5.1

Protection

56
■■■■■■■■
The supply bottle with the eluent must be connected as indicated in chap-
ter 3.8.1, page 24. This is above all important for eluents with volatile sol-
vents (e.g. acetone).
Moreover, condensation must also be prevented in the eluent bottle. Drop
formation can change the concentration ratio in the eluent.
In the case of very sensitive measurements, we recommend that the elu-
ent be stirred constantly with a magnetic stirrer (e. g. the 2.801.0010 with
6.2070.000).
To protect the IC system against foreign particles, we recommend aspirat-
ing the eluents via a 6.2821.090 aspiration filter (9-2). This aspiration filter
must be replaced should it show signs of yellow discoloration (but no later
than every 3 months).
When changing the eluent, it must be ensured that no precipitates can
occur. Solutions following one another in direct succession must therefore
be miscible. If the system has to be rinsed organically, several solvents
with rising or falling lipophilia must be used.
Caution
The pump head is filled ex works with methanol/ultrapure water. It
must be ensured that the eluent used is freely miscible with the solvent
remaining in the pump head.
To protect the high pressure pump against foreign particles, we recom-
mend that the eluent undergoes a microfiltration (filter 0.45 µm) before
being aspirated via a 6.2821.090 aspiration filter (see Chapter 5.4.2.2,
page 56).
Salt crystals between the piston and seal cause abrasion particles which
can find their way into the eluent. These lead to contaminated valves, a
rise in pressure and in extreme cases scratched pistons. It is therefore
essential to ensure that no precipitates can occur (see Chapter 5.4.2.3,
page 56).
■■■■■■■■■■■■■■■■■■■■■■
882 Compact IC plus – Anion

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents