3.1 About this chapter

3 Installation

3.1
About this chapter
3.2

Initial installation

10
■■■■■■■■
The Installation chapter contains:
this overview.
a brief set of instructions for the initial installation of the 882 Compact
IC plus – Anion. At each step you will find cross-references to more
detailed installation instructions for individual components, should you
require such aids.
an installation diagram (see Figure 3, page 14), showing a com-
pletely installed 882 Compact IC plus – Anion.
several chapters (Chapter 3.4, page 15 and following) with detailed
installation instructions for all components, including those that are
already installed at the time the instrument is delivered.
Note
A number of the capillaries is already connected at the time the instru-
ment is delivered.
Installing 882 Compact IC plus – Anion
Install the instrument as follows:
1 Setting up the instrument
(see Chapter 3.4, page 15).
Place the instrument in a location of the laboratory which is suita-
ble for operation and free of vibrations.
The installation site must be protected against corrosive atmos-
phere and contamination by chemicals. If possible, avoid excessive
temperature fluctuations and direct sunlight.
2 Installations on the rear of the instrument
Place the detector in the instrument and connect it (see manual
of the detector).
Remove transport locking screws (see Chapter 3.6.1, page 19).
■■■■■■■■■■■■■■■■■■■■■■
882 Compact IC plus – Anion

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 882 Compact IC plus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

882 compact ic plus anion2.882.00201.882.0020881 compact ic pro series881 compact ic pro anion

Table of Contents