3.1 About this chapter
3 Installation
3.1
About this chapter
3.2
Initial installation
10
■■■■■■■■
The Installation chapter contains:
this overview.
■
a brief set of instructions for the initial installation of the 882 Compact
■
IC plus – Anion. At each step you will find cross-references to more
detailed installation instructions for individual components, should you
require such aids.
an installation diagram (see Figure 3, page 14), showing a com-
■
pletely installed 882 Compact IC plus – Anion.
several chapters (Chapter 3.4, page 15 and following) with detailed
■
installation instructions for all components, including those that are
already installed at the time the instrument is delivered.
Note
A number of the capillaries is already connected at the time the instru-
ment is delivered.
Installing 882 Compact IC plus – Anion
Install the instrument as follows:
1 Setting up the instrument
(see Chapter 3.4, page 15).
Place the instrument in a location of the laboratory which is suita-
■
ble for operation and free of vibrations.
The installation site must be protected against corrosive atmos-
phere and contamination by chemicals. If possible, avoid excessive
temperature fluctuations and direct sunlight.
2 Installations on the rear of the instrument
Place the detector in the instrument and connect it (see manual
■
of the detector).
Remove transport locking screws (see Chapter 3.6.1, page 19).
■
■■■■■■■■■■■■■■■■■■■■■■