Current Mode Operation; Front Panel Function Checks - Xantrex XKW 8-350 Operating Manual

Xkw 3000 watt series programmable dc power supply
Table of Contents

Advertisement

Installation
Functional Tests
Current Mode
To check current mode operation, proceed as follows:
Operation
1. Rotate the VOLTAGE and CURRENT controls fully counterclockwise.
2. Rotate the VOLTAGE control 1/2 turn clockwise.
3. Connect a high current DC ammeter
4. Set the POWER switch to ON.
5. Rotate the CURRENT control slowly clockwise. The control range should be
6. Set the POWER switch to OFF.
Front Panel
1. Press OVP CHECK switch and check that the voltmeter displays approximately
Function
Checks
2. Turn OVP SET potentiometer counter-clockwise and check that the voltmeter
3. With voltage and current controls turned all the way in a clockwise direction,
4. With voltage and current controls turned all the way in a clockwise direction,
5. Push REM/LOC switch to IN position and check that the REM LED illuminates
1.
36
correct polarity. Select leads of sufficient current carrying capacity and an
ammeter range compatible with the unit's rated current output. The ammeter
should have an accuracy of better than 0.5%.
from zero to the maximum rated output. Compare the test meter reading with the
reading on the front panel ammeter. Check that the red current mode indicator
led is ON.
the model-rated output voltage plus 10%.
reading decreases. Continued turning (up to 20 turns) will see the reading
decrease to approximately 5% of the model-rated voltage output. Turn the
potentiometer clockwise until the voltmeter once again displays approximately
the model-rated output voltage plus 10%.
press the V & I CHECK switch and check that the voltmeter and ammeter
display a minimum of the power supply model output ratings.
push the STANDBY switch to its IN position and check that the voltmeter
reading falls to zero and the S/D (Shutdown) LED illuminates. Push the
STANDBY switch once again to reset it to its OUT position. The S/D LED will
turn off.
and the voltmeter reading falls to zero. Reset the REM/LOC switch to its OUT
position for local (default) operation.
Either a direct reading meter or calibrated meter and shunt combination.
1
across the rear output terminals, observing
Operating Manual for XKW 3kW Series Power Supply