Software Configuration Checks

Press the UTILITIES key and select the SYSTEM item from the "Utilities" menu - which is shown in the
menu area on the left side of the screen. The System Setup desktop offers different pages to check:
Table 4-10
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
Press the UTILITIES key and select the MEASURE item from the "Utilities" menu - which is shown in
the menu area on the left side of the screen.
The Measure Setup desktop offers three different pages to check.
Parameters and possible adjustments depend on the selected Application.
NOTICE
To view, add, delete, reorder, edit or when creating a new parameter (in the
is very important that all items are chosen correctly and that the relevant item is highlighted.
For further information refer to Chapter 18 in the Basic User Manual of Voluson® 730Pro / 730ProV.
Table 4-11
Step
11
2
3
RETZTECHNIK
2
EVISION

Software Configuration (System Setup) Checks

Task
General: Check Date and Time setting
General: Check that Location (Clinic Name) is correct
General: Check Language settings
User Setting: Check all the User Settings
User Settings: Check settings of the "Patient ID" field
Peripherals: Check assignment of Printer keys
Peripherals: Check assignment of Foot switch
Peripherals: Check Save Destination assignment
Peripherals: Check 2D Save key assignment
Peripherals: Check Remote Print A, Remote Print B,
Report Printer and DICOM Print Job assignment
Option: Check that all options are set up correct
Service: Check the Video Norm standard as described in
Section 3-8-3-4 on page 3-42
Network:
Check DICOM, Sonoview and Network configuration

Measurement Setup Checks

Task
Measure & Calc:
Check all settings for all applications
Application Parameters:
Check all settings for all applications
Global Parameters: Check all settings
Chapter 4 - Functional Checks
DRAFT
V
OLUSON
Date and Time are correct
Location Name is correct
desired Language is displayed
settings assigned as desired by the customer
settings assigned as desired by the customer
Print A and Print B keys are assigned as desired by the
customer
Foot Switch left and right are assigned as desired by the
customer
assigned as desired by the customer
assigned as desired by the customer
settings assigned as desired by the customer
D = Demo , I = Inactive , P = Permanent
settings assigned as required for the country
settings assigned as desired by the customer
setting assigned as desired by the customer
setting assigned as desired by the customer
setting assigned as desired by the customer
® 730P
/ 730P
RO
RO
S
ERVICE
Expected Result(s)
MEASURE & CALC
Expected Result(s)
V (BT´04)
M
ANUAL
page), it
4-41