Section 8-10
Replacement of the Monitor Task Lamp
8-10-1
Manpower
One person, 15 min.
8-10-2
Tools
Philips screwdriver 1 and 2, stubby (length ~30mm)
8-10-3
Preparations
1.) Power Off/Shutdown the system; see:
8-10-4
Task Lamp - Removal Procedure
1.) Use the stubby screwdriver and unscrew the 2 screws, which fixes the cover of the task lamp.
2.) Screw out the task lamp and remove it.
8-10-5
Task Lamp - Installation Procedure
1.) Screw in the new task lamp into the lamp socket.
2.) Mount the cover of the task lamp and fix it with the 2 screws.
8-22
RETZTECHNIK
2
EVISION
Figure 8-28 remove cover of the task lamp
Section 8-10 - Replacement of the Monitor Task Lamp