Site Circuit Breaker

It is recommended that the branch circuit breaker for the machine be readily accessible.
POWER OUTAGE MAY OCCUR.
CAUTION
The Voluson® 730Pro / 730ProV requires a dedicated single branch circuit. To avoid circuit
overload and possible loss of critical care equipment, make sure you DO NOT have any other
equipment operating on the same circuit.
2-2-2-4

Site Power Outlets

A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5

Unit Power Plug

If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.