Trackball Adjustment

Adjustment of the mechanical movement may be necessary to ensure smooth running of the trackball.
1.) Remove the outer fixing ring by turning it counterclockwise.
rotate clockwise to
mount the fixing ring
2.) Adjust the trackball for smooth running by rotating the black securing ring.
CCW: smooth run
CW: tighten run
tighten run
Avoid tightening of thread caused by improper mounting of securing ring!
NOTICE
3.) Mount the outer fixing ring by turning it clockwise. see
6-6