Pressure transmitter with 4-20 ma hart revision 5 and 7 selectable protocol
Hide thumbs Also See for Rosemount 3051:
Table of Contents

Advertisement

Troubleshooting
August 2019
2. Using the two captive screws as handles, insert the electronics board into the
3. Tighten the captive mounting screws.
4.
6.6.2
Install terminal block
Complete the following steps to reinstall the terminal block into the transmitter.
Procedure
1.
2. Tighten the captive screws.
3.
112
housing.
Make sure the power posts from the electronics housing properly engage the
receptacles on the electronics board. Do not force. The electronics board should
slide gently on the connections.
WARNING
Explosions
Explosions could result in death or serious injury.
In an explosion-proof/flameproof installation, do not remove the transmitter
covers when power is applied to the unit.
The transmitter covers must be engagged metal-to-metal to ensure a proper
seal and to meet explosion-proof requirements.
Replace the electronics housing cover.
WARNING
Electrical shock
Electrical shock can result in death or serious injury.
Avoid contact with the leads and terminals. High voltage that may be present on
leads can cause electrical shock.
Gently slide the terminal block into place, making sure the two power posts from
the electronics housing properly engage the receptacles on the terminal block.
WARNING
Explosions
Explosions could result in death or serious injury.
The transmitter covers must be fully engaged to meet Explosion-Proof
requirements.
Replace the electronics housing cover.
Reference Manual
00809-0100-4007
Rosemount 3051

Hide quick links: