Fault – Cause – Remedy
Message
!
O
supply pressure high
2
!!!
PEEP high
!!!
PEEP valve inop.
!
Pressure limited
!!!
Pressure meas. inop.
!!!
Standby activated
!!!
Temperature high
!!!
Temperature meas. inop.
!!!
Temperature sensor ?
!!!
Tidal volume high
!
Tidal volume high
!!!
Tube blokked
!!
Volume not constant
!!!
CO
measurement inop
2
88
Cause
O
supply pressure too high.
2
O
supply pressure is not required when
2
FiO
= 21 Vol.%.
2
Expiratory system obstructed.
Expiratory resistance is increasing.
Machine faulty.
Internal PEEP valve faulty.
P
pressure limit is active.
max
Fluid in expiration valve.
Pressure measurement malfunction.
Evita 2 dura has been switched to
standby.
Breathing gas temperature higher than
o
40
C.
Temperature sensor faulty.
Temperature sensor probe has been
disconnected during operation.
Sensor cable broken.
The upper alarm limit of the applied
inspiratory tidal volume V
has been
T
exceeded during three consecutive
ventilation strokes.
Leak or disconnection.
The inspiratory tidal volume V
exceeded the upper alarm limit.
Leak or disconnection.
Evita 2 dura only applies a very small
volume with each mechanical stroke,
e.g. because the tube is blocked.
Patient "fights" against the mechanical
strokes in pressure-controlled ventilation,
so that the set inspiratory pressure volume
is achieved with only a very small volume.
Due to pressure limit or time limit, the set
tidal volume V
has not been applied.
T
CO
sensor faulty.
2
CO
electronics in unit faulty.
2
Remedy
Make sure pressure is less than 6 bar.
Check hose system and expiration valve.
Check bacterial filter. Replace if
necessary.
Call DrägerService.
Call DrägerService.
Check condition of patient, check pattern
of ventilation, correct setting if necessary.
Replace expiration valve, page 92, then
clean and dry.
Call DrägerService.
Confirm standby with »Alarm Reset« key.
Switch off humidifier.
Fit new temperature sensor, see page 22.
Reconnect probe.
Fit new temperature sensor.
Check condition of patient,check pattern
of ventilation, correct alarm limit if
necessary.
Check that hose system connections are
leakproof.
has
Check condition of patient,check pattern
T
of ventilation, correct alarm limit if
necessary.
Check that hose system connections are
leakproof.
Check condition of patient, check tube.
Check condition of patient, check
machine settings.
Prolong inspiratory time »T
Increase inspiratory flow »Flow«
Increase pressure limit »P
Press the »Alarm Reset« key to suppress
the visual and acoustic alarm until the
cause of the alarm is remedied.
Replace defective CO
Call Dräger Service.
«
insp
«.
max
sensor.
2