Insp. O 2 Concentration During Medicament Nebulisation - Dräger Evita 2 dura Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Evita 2 dura:
Table of Contents

Advertisement

Insp. O
concentration during medicament nebulisation
2
Use only medicament nebuliser 84 12 935 (white central
body).
If other medicament nebulisers are used, considerable
deviations may occur in the tidal volume and inspiratory
O
concentration!
2
To minimise the deviation from the set O
Evita 2 dura generates a mixed gas to drive the medicament
nebuliser.
In adult ventilation, this mixed gas is generated by switching
over the compressed gases (medical air and oxygen) in
synchronisation with inspiration.
In paediatric ventilation, the nebuliser is operated
continuously, with medical air or oxygen in alternation. The
drive gas of the medicament nebuliser therefore roughly
corresponds to the set FiO
The graph shows the possible deviations of the applied
O
concentration as a function of the set FiO
2
inspiratory flow (15 L/min) in adult ventilation, or at ventilation
frequencies above 12 bpm in paediatric ventilation.
concentration,
2
.
2
at a minimal
2
100
90
80
70
60
50
40
30
20
20
30
40
50
set O
concentration FiO
2
Description
Measurements
60
70
80
90
100
Vol.%
2
127