Maintenance Intervals - Dräger Evita 2 dura Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Evita 2 dura:
Table of Contents

Advertisement

Maintenance Intervals

Clean and disinfect equipment and/or components before
any maintenance procedures and/or before returning for
repair!
Note the maintenance intervals specified in the respective
Instructions for Use of the options installed!
O
sensor capsule
2
Ambient-air filter
Cooling-air filter
Filters in the compressed gas inlets To be replaced by trained service personnel
Lithium battery for data protection
Real-time clock
Pressure reducer
Equipment inspection and service
Replace when the following message is
displayed:
»O
measurement inop«
2
and when calibration is no longer possible.
Spent O
sensors can be returned to Dräger
2
for disposal.
Clean or replace after 4 weeks, see page 98.
Replace after 1 year.
Dispose of with normal domestic waste.
every 2 years.
To be replaced by trained service personnel
every 2 years.
Disposal: see page 99.
To be replaced by trained service personnel
every 6 years.
Disposal: page 99.
To be replaced every 6 years by
DrägerService.
Every 6 months by trained service personnel.
Care
Maintenance Intervals
97