Operation
Special Functions
1 Connect the nebuliser hose to the port on the front panel of
the Evita 2 dura.
Fill the medicament nebuliser in accordance with its
specific Instructions for Use.
Warning: the effect of aerosols on sensors, filters and
heat and moisture exchangers (HME) must be taken into
account!
The measuring function of the flow sensor may be impaired.
The flow resistance of filters is liable to increase and may
impair ventilation.
Do not place a microbial filter on the nebuliser outlet
during nebulisation!
During medicament nebulisation, do not use a heat and
moisture exchanger (HME) at the Y-piece. Risk of
increased breathing resistance!
2 Hold down key »
« until the yellow LED lights up.
Advisory message on the screen:
»Nebuliser on!«
The nebuliser remains in operation for 30 minutes.
If you wish to interrupt medicament nebulisation before it is
complete:
2 Press »
« again. The yellow LED goes out, the
nebuliser will be switched off.
The flow sensor is subsequently cleaned and calibrated
automatically.
Remove residual medicament. Strictly follow the
Instructions for Use of the medicament nebuliser.
62
D
1
D
2
Evita 2 dura
Evita 2 dura