Disposable Biopsy Needle Type A

Disposable Biopsy Needle Type A

Navigating the Disposable Biopsy Needle

Overview
Once the disposable biopsy needle has been correctly assembled with the
page 188) or the
the planned target point.
How the Needle is
The insertion depth of the needle is tracked by the two reflective discs on the needle's track-
Tracked
ing array.
The trajectory of the needle is tracked by:
• The
• The
Both reflective discs of the needle's tracking array must be in the camera's direct field of
Tracking Guidelines
view.
Position the needle's tracking array at a right angle to the camera and in the center of the
camera volume.
The
all times.
200
Frameless Biopsy System
VarioGuide
tracking array if you are using the
Biopsy Alignment Array
VarioGuide
tracking array or
The
Pointer
is not tracked while the
field of view. It can be tracked again if the
from the camera field of view for at least 2 seconds.
(see page 198), you can navigate the needle to
VarioGuide
if you are using the
Frameless Biopsy System
Biopsy Alignment Array
Disposable Biopsy Needle
Disposable Biopsy Needle
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
VarioGuide
must be visible to the camera at
is in the camera
is hidden
(see