Preparation For Registration - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Introduction

Preparation for Registration

Reference Arrays
To enable registration and navigation, attach a reference array (e.g., the
Array
tient's head throughout the procedure.
The reference array must be visible to the camera at all times.
Camera Positioning
When you select a registration method, the software displays the following dialog if the ref-
erence array is not visible to the camera:
Step
Adjust camera distance and angle so that the reference array (indicated by colored rings
and spheres) is represented in the center of the blue cone. The rings should be as close to
the center as possible, so that there is sufficient space for navigation instruments.
NOTE: For information on camera positioning, see page 48.
Sound Output
Activating the
58
) to the patient. This allows the system to track the location and movement of the pa-
Sound Output
Beep
Explanation
High pitched
Reference array is close to the center of the camera field of view
Low pitched
Reference array is further from the center of the camera field of view
Figure 14
button provides acoustic guidance during positioning.
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
Mayfield Reference