Configuring Microscope Video - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Microscope Tools

Configuring Microscope Video

How to Configure a
Video View
Steps
218
1.
Press
Video Configuration
2.
In the
Input
tab, press
panel the microscope video is connected).
By default,
Input 1
User Guide
.
3.
In the
Settings
tab, adjust image view settings as needed using the slider bars.
NOTE: Video image settings are described on page 52.
4.
Press
to confirm settings.
Accept
The installation and calibration of the video hardware must be performed by
qualified BrainLAB personnel. Removing and remounting or modifying the
hardware from the microscope or replacing the complete microscope will sig-
nificantly reduce the accuracy of the contours displayed in the video overlay
window or image injection unit. Thus, the calibration of the video overlay win-
dow must be renewed in these cases.
During each intervention, make sure that the video is connected to the same in-
put selected during calibration. Using a different input may result in an incor-
rect calibration.
Figure 96
in the
Tools > Microscope
Input 1
or
Input 2
(depending on to which
is linked to the microscope video input. See also the
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
dialog.
Kolibri
user
System