Calibration And Verification Using Pivot Points - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Calibration Using the ICM4

Calibration and Verification Using Pivot Points

How to Calibrate
and Verify Using
Pivot Points
Steps
Other Options
You can define an instrument diameter by pressing
ibration
142
1.
Hold
ICM4
and instrument with
of view.
The
Instrument Calibration
2.
Insert instrument tip into a pivot point and rotate instrument slightly.
3.
Once tip calibration is complete, the
4.
Insert instrument tip into same pivot point used for calibration and slightly rotate
instrument to display the tip deviation.
5.
To confirm calibration, press
6.
Remove
from camera field of view.
ICM4
dialog (see page 140).
Figure 56
StarLink
tracking array attached in camera field
dialog opens.
Verification of Calibration
, or press
Accept
Try Again
Set Diameter
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
dialog opens.
to repeat calibration.
in the
Verification of Cal-