Calibration Using the Reference Array
Calibration Using the Reference Array
Introduction
About the
Reference Array
You can perform instrument calibration and verification using the reference array cone a.
Suitable
Instruments
Reference Array
Position
138
Suitable
Instruments with a pointed tip
Instruments with:
• A round tip
• A large diameter that does not fit correctly into the cone of the reference
Not
array
suitable
NOTE: Curved instruments can be accurately calibrated, however, the
axis of these instruments remains uncalibrated. To calibrate curved instru-
ments or to gain information about the diameter of the instrument, use the
ICM4 or ICM 2.0 .
Make sure that the reference array position does not change during calibration
verification.
a
Figure 52
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7