Verifying Instruments - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Frameless Biopsy System

Verifying Instruments

How to Verify
Disposable Biopsy
Needle
Steps
192
1.
Open
Tools > Tool Calibration/Verification
The
Biopsy Needle Verification
2.
Insert
Disposable Biopsy Needle Type A
ment Array
attached until
The arrows indicate the direction in which to adjust the needle until the insertion
depth is achieved.
3.
Secure needle in biopsy tube by tightening fixator screw on tube.
4.
Bring biopsy tube and needle into a vertical position.
5.
The software calculates and displays the trajectory deviation of the biopsy needle
based on the coordinates of the tracking array on the needle and the
.
Alignment Array
6.
Press
to return to the
Close
Figure 85
and press
dialog opens.
into biopsy tube with
Stop
is highlighted in blue.
Tool Calibration/Verification
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
Biopsy Needle
.
Biopsy Align-
Biopsy
dialog.