Microscope Navigation

Verifying the Microscope

Introduction

Overview
The microscope is calibrated to work as a pointer with its tip at the focal plane. You can ver-
ify the calibration by focusing on a landmark in the center of the field of view and comparing
this to the data shown in the navigation view.
NOTE: Always verify the accuracy of the microscope calibration.
The Verification
Dialogs
No.
a
s
d
f
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
a
Component
Microscope Depthview
: Displays a reconstruction of the microscope's field of
view (see page 208).
Microscope video view (if available, see page 220)
Distance from microscope to center of reference array (cone)
Video Verification
button: Press to toggle between the microscope views and ax-
ial, sagittal and coronal views
s
Figure 90
d
f
207