Introduction to Calibration
Handling Instrument Adapters
Attachment of
To help ensure successful calibration, make sure that:
Instrument
• You have attached an adapter that is large enough for the instrument. The larger the
Adapters
• The long arm of the adapter's tracking array runs parallel to the long axis of the instru-
Unsuccessful
If the software recognizes that an instrument adapter is too small for an instrument, the di-
Calibration
alog below is displayed.
You can now attach a larger adapter, or change the adapter position on the instrument and
press
132
instrument, the larger the adapter you should use. Using an adapter that is too small can
result in imprecise calibration.
ment to which it is attached, as shown in the image below.
No.
Component
Long arm of tracking array
a
Long instrument axis
s
to repeat calibration.
Try Again
a
Figure 47
Figure 48
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7