Performing Calibration - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Calibrating Instruments

Performing Calibration

How to Calibrate
and Verify
Steps
1.
2.
3.
4.
5.
Instrument
Trajectory
After calibrating an instrument using the cone, the plane of the instrument
adapter array is orthogonally transferred until it intersects with the instrument
tip. Thus, the calibrated instrument trajectory may not correspond to the actual
instrument axis.
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
Hold instrument with
StarLink
If you are calibrating the instrument for the first time, the
dialog opens.
NOTE: If the instrument has already been calibrated in this session, proceed to
step 4.
Hold instrument tip to reference array cone and slightly pivot instrument.
Once calibration is complete, the
Pivot instrument in reference array cone to display tip deviation.
To confirm calibration, press
tion.
Figure 53
tracking array attached in camera field of view.
Instrument Calibration
Verification of Calibration
, or press
Accept
Recalibration
dialog opens.
to repeat calibra-
139