Verifying Instruments - Brainlab Kolibri cranial Software User's Manual

Navigation system
Table of Contents

Advertisement

Navigating Alignment Instruments

Verifying Instruments

How to Verify the
VarioGuide
Steps
1.
2.
3.
Troubleshooting
If the accuracy is insufficient, make sure that:
VarioGuide
• The
Verification
• The guide disk is properly attached
• The
Software User Guide Rev. 1.0 Kolibri cranial/ENT Ver. 2.7
a
Open
Tools > Tool Calibration/Verification
The
VarioGuide Verification
Place sharp tip pointer in pivot points a one after the other on guide disk. The
software displays the deviation. The guide disk contains eight pivot points. You
should verify at least three points, depending on accessibility due to the
VarioGuide
position.
When the overall deviation is less than 1.0 mm, the light s is shown green indi-
cating that the accuracy is sufficient.
Press
to return to the
Close
is correctly assembled
VarioGuide
VarioGuide
array is not bent
Figure 78
and press
VarioGuide
dialog opens.
Tool Calibration/Verification
s
.
dialog.
179