Advertisement

Table 60: DC test failure types
Status
Failure type
FAIL
Over voltage
FAIL
Open ground
FAIL
Min Meg-Ohm
FAIL
Min PI
FAIL
Over current
FAIL
No steps defined
FAIL
Low PRG HPT voltage The last step voltage is below the previous Meg-Ohm/PI test
Table 61: Surge test failure types
Status
Failure type
FAIL
Open leads
FAIL
Over voltage
FAIL
Open ground
FAIL
ppEAR Limit
FAIL
L-L EAR
CAUTION
Unequal zero
crossings
168
Baker AWA-IV 71-015 EN V16 User Guide
Failure message
Over Voltage Condition has been detected.
An open ground has been detected. Check power cord outlet
ground for continuity.
Test Result. Fail – MIN MEG-OHM. Meg-Ohm value is less than
the minimum tolerance.
PI test results: Fail – Min PI. PI Ratio is less than minimum
tolerance.
Test result: FAIL – OVER CURRENT. An over current trip was
detected.
No steps for Step-Voltage Test exist in this Test ID.
voltage.
Failure message
Test Results: Fail – OPEN LEADS. Open leads detected.
Over Voltage Condition has been detected.
An open ground has been detected. Check power cord power
outlet ground for continuity.
Surge test results: Fail – ppEAR LIMIT. Pulse-to-Pulse EAR%.
Surge Test Result: Fail – L-L EAR Lead- to-Lead EAR % is out of
tolerance.
Test results: Caution – UNEQUAL ZERO CROSSINGS Waveforms
do not have the same number of zero crossings.