AC amplitude (high impedance) verification

This test checks AC amplitude output accuracy at 1 kHz frequency using each attenuator.
1. Set the multimeter to measure Vrms. Connect the multimeter to the channel output as shown above.
2. Set the instrument to each output in the table below and measure the output voltage with the multimeter. Be
sure the output impedance is set to High–Z and the output is enabled.
Output Setup
Function
[b]
High Z
Sine
High Z
Sine
High Z
Sine
High Z
Sine
High Z
Sine
[a] Based upon 2% of setting ±1 mVpp; converted to Vrms for High–Z.
[b] Use the following sequence to set this output:
1. Set amplitude to 400.0 mVrms
2. Set DC Offset to 1.0 VDC
3. Set Auto-Range to OFF
4. Set DC Offset Voltage to 0.0 VDC
5. After the measurement, set autorange ON for remaining measurements.
3. Compare the measured value to the test limits shown in the table.
4. For EDU33212A only. Connect multimeter to channel 2 output and repeat steps 2 and 3.
18
Waveform Generator
Amplitude
400 mVrms
400 mVrms
1 Vrms
2.5 Vrms
7 Vrms
Measurement
Frequency
Nominal
1 kHz
400 mVrms
1 kHz
400 mVrms
1 kHz
1 Vrms
1 kHz
2.5 Vrms
1 kHz
7 Vrms
Keysight EDU33210 Series Service Guide
[a]
Error
±0.008354 Vrms
±0.008354 Vrms
±0.020354 Vrms
±0.050354 Vrms
±0.140354 Vrms