9) Position the lamination anchor in the longitudinal residual limb
axis on the hemispherical dummy.
10) Bring the lamination anchor arms to the correct a-p and m-l posi­
tion. Never align the threaded clamping in the anterior or posteri­
or direction.
11) Optional: Adapt the anchor arms of the lamination anchor to the
model with a bending iron.
12) Replace the shuttle lock with the lamination dummy (see
Page 12).
13) Wrap the clamping section of the lamination anchor with plasta
tape to prevent contact with the lamination resin. This ensures
even clamping.
14) Pull the fibreglass roving through the holes in the anchor arms
and allow it to hang in a loop. If there are no holes, allow the
fibreglass roving to hang around the anchor arms in a loop.
15) Place 3 layers of carbon fibre cloth (e.g. 15 cm x 15 cm) offset
over the lamination anchor arms.
16) Pull 2 layers of Perlon stockinette over the model and tie off below
the cap screw of the lamination anchor.
17) Pull 1 layer of carbon fibre woven hose over the model.
18) Cut 1 piece of carbon fibre woven hose to size (1.3 times the
length of the plaster impression) and pull it over the model.
19) Tie off the excess carbon fibre woven hose below the cap screw
of the lamination anchor and pull it over the plaster impression.
20) Pull 2 layers of Perlon stockinette over the model and tie off below
the cap screw of the lamination anchor.
21) Pull on a soaked PVA bag.
22) Complete the lamination process with Orthocryl lamination resin.
23) After the lamination resin has cured, remove the lamination
dummy.
5.2.3 Installing the lamination dummy
1) Tighten the screw of the threaded clamping to a point where the
shuttle lock can still be screwed in and out.
12
2) Unscrew the shuttle lock from the lamination anchor.
3) Screw in the lamination dummy.
5.3 Final assembly
Installing the shuttle lock
Required materials: 710D4 torque wrench, 636K13 Loctite® 241
>
1) Screw the shuttle lock completely into the thread of the lamination
anchor.
2) The shuttle lock may be unscrewed from the thread by no more
than 3/4 of a turn to align it.
3) Secure the cap screw of the threaded clamping using Loctite®
and tighten using the torque wrench (10 Nm).
Screwing in the ratchet unit
>
Required materials: 710D4 torque wrench
► Screw the ratchet unit into the shuttle lock (torque value: 10 Nm).
6 Use
CAUTION
Installation of the pin in a liner that is not approved
Risk of injury due to loosening of the screw connection
► Only install the pin in liners with a metal thread and observe the
specified tightening torque.
CAUTION
Pin not fully inserted into the lock
Risk of injury due to loss of connection to the prosthesis
► Guide the pin into the lock in accordance with the instructions.
► Always verify that the pin is engaged in the lock before using the
prosthesis.
The liner and shuttle lock are connected to each other with the pin.
The pin engages in the shuttle lock and holds the liner in place.

Advertisement