3-1. Troubleshooting
3-1-1. MKS-8010A
The main power cannot be turned on.
(Indicator does not light green.)
Flow 1
The main power of the
MKS-8010A cannot be turned on.
(Indicator does not light green.)
Is the POWER switch of the
MKS-8010A turned ON?
YES
Is the power cord of the
MKS-8010A connected correctly?
YES
The power supply unit is probably
defective.
CCP-8000 IM
Section 3
Service Overview
NO
Set it to ON.
NO
Connect it
correctly.
3-1-2. MKS-8011A
Nothing is displayed on the menu monitor screen.
Flow 2
Nothing is displayed on the menu
monitor screen.
Is the POWER switch of the
MKS-8010A turned ON?
YES
Is the connection cable is connected
correctly? (Is the connector (SCU)
of the MKS-8011A connected to
the connector (MENU PANEL) of
the MKS-8010A?)
YES
Is the switch on the MKS-8011A
lit?
YES
1. The backlight of the MKS-8011A
is probably burnt out.
2. LCD or the LCD drive circuit of
the MKS-8011A is probably
defective.
Check the Flow 1
NO
diagram "The main
power cannot be
turned on."
NO
Connect the D-sub
50-pin cable.
NO
The MKS-8011A is
probably defective.