Status LED Charging status indicat­
The
LED
ring is lit up
green
The
LED
ring is flash­
ing red

18.2.3 Error messages while establishing a connection with the cockpit app

Error message
Device not found
Component was connec­
ted
to
another
device.
Establish connection?

18.2.4 Status signals

Battery charger connected
LED on
Status LED
power sup­
on battery
ply
charger
Battery charger disconnected
Beep sig­
Vibration
nal
signal
1x short
1x short
Genium X3 3B5-2/3B5-2=ST
or (5 LEDs)
Cause
The connection could not be
established as no component
with the entered serial num­
ber could be found.
The component was connec­
ted to another device/remote
control.
A current connection to the
prosthesis was interrupted.
Event
Power supply and battery charger operational. Battery charger not yet connected to
receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to avoid bother­
some light at night. The charging process is not affected.
Event
Self-test completed successfully. Product
is operational.
Error
The
battery
charger
functional but not connec­
ted to the receiver, or the
distance between the bat­
tery charger and receiver
of the charging unit is too
great.
The prosthesis is not being
charged
Defective battery charger.
Correction
Compare the serial number entered with that of the
component, then try to establish the connection
again.
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
Check the following points:
Distance from the prosthesis to the device
Battery charge level of the prosthesis
Bluetooth on the prosthesis switched on? (see
Page 37)
If multiple prostheses were stored in the remote
control, was the correct prosthesis selected?
Resolution
Appendices
Resolution
is
Connect the battery charger
or
reduce
the
distance
between the charger and
the receiver of the charging
unit on the prosthesis.
Reset the error by discon­
necting and connecting the
power supply.
If the error persists, the bat­
tery charger and power sup­
ply must be inspected by an
authorised Ottobock Ser­
vice Center.