Operating States/Error Signals; Signals For Operating States; Warnings/Error Signals - Otto Bock Genium X3 3B5-2 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-2:
Table of Contents

Advertisement

Check
Ottobock Data Station.

18.2 Operating states/error signals

The prosthesis indicates operating states and error messages through beeps and vibration signals.

18.2.1 Signals for operating states

Battery charger connected/disconnected
Beep signal
1x short
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app or using the remote control, there are no beep signals
(see Page 34).
Beep signal
Vibration sig­
nal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short

18.2.2 Warnings/error signals

Error during use
Beep signal
5x long
Genium X3 3B5-2/3B5-2=ST
the
product
settings
Vibration signal
3x long
1x before beep signal
Additional action performed
Mode switching using the Cockpit
app or remote control
Bouncing on the forefoot followed
by unloading the prosthetic leg
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Vibration signal
1x long at intervals of
approx.
5 seconds
mute mode (silent mode)
is activated, this signal is
not generated)
3x long
5x long
5 x long repeated every
60 seconds
using
the
corresponding
Event
Charging mode started (3 sec. after connecting the bat­
tery charger)
Self-test completed successfully, product is operational
Event
Mode switching performed using the
Cockpit app or remote control.
Bouncing pattern recognised.
Switching to basic mode (mode 1) car­
ried out.
Switching to MyMode 1 (mode 2) car­
ried out.
Switching to MyMode 2 (mode 3) car­
ried out.
Event
Overheated hydraulic unit Reduce activity.
(if
Battery
charge
under 25%
Battery
charge
under 10%
Moderately
severe
error/indication
safety mode activation
e.g. a sensor is not oper­
ational
Switch to safety mode
(see Page 43).
adjustment
software
Required action
level
Charge battery soon.
Remaining
operating
approx. 24 hours
level
Charge battery soon
Remaining
operating
approx. 6 hours
Restricted walking possible.
of
The change in flexion resist­
ance has to be noted.
The product must be inspec­
ted immediately by an author­
ised Ottobock Service Center.
Appendices
of
the
time
time
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Genium x3 3b5-2-st

Table of Contents