Use
Parameter
Stair Function
Pitch
Volume
In addition to the parameters, the remote control (optional accessory) can also be used to activate deep sleep
mode. Further information on deep sleep mode (see Page 38).

10.2.3 Overview of adjustment parameters in MyModes

INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific motion pattern such
as cross-country skiing. Damping behaviour is not automatically controlled and adjusted in MyModes.
The following parameters can be modified in the MyModes (except MyMode "Brake knee"):
Parameter
Gain
Pitch
Volume
The following parameters can be modified in MyMode "Brake knee":
Parameter
Stance function
Pitch
Volume
36
Adjustment soft­
App/remote
ware range
control adjust­
ment range
Deactivated
0 - deactivated
Activated
1 - activated
1000 Hz
1000 Hz
4000 Hz
4000 Hz
0 – 4
0 – 4
Adjustment soft­
App/remote
ware range
control adjust­
ment range
0 – 100
+/- 10
1000 Hz
1000 Hz
4000 Hz
4000 Hz
0 – 4
0 – 4
Adjustment soft­
App/remote
ware range
control adjust­
ment range
0 – 180
+/- 20
1000 Hz
1000 Hz
4000 Hz
4000 Hz
0 – 4
0 – 4
Meaning
Activating/deactivating the stair and obstacle
function. For switching with the Cockpit app
or remote control, this function has to be
activated in the adjustment software. Further
information (see Page 32).
Pitch of beep signal for confirmation tones
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals in case of errors are still gen­
erated.
Meaning
Value for speed at which flexion resistance
increases as the knee angle increases
Pitch of beep signal for confirmation tones
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals in case of errors are still gen­
erated.
Meaning
Basic resistance for the knee joint without
load
Pitch of beep signal for confirmation tones
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals in case of errors are still gen­
erated.