Beep sig­
Vibration
nal
signal
3x short
3x short
5x long
5x
long
(every
minute)
Battery charge level
During the charging process, the current battery charge level is indicated by the LEDs lit on the side of the char­
ger.
LEDs
0
Battery charge
0%-10%
level
52
Event
Maintenance note:
e.g. service interval has been exceeded,
temporary disruption of a sensor signal
Self-test error
AXON tube adapter not connected while
removing inductive charger
1
2
11%-30%
31%-50%
Resolution
Use the Cockpit app/remote control
to check the next maintenance date
for the prosthesis (see Page 37). If
the
date
has
exceeded, the prosthesis, the battery
charger, the power supply and the
remote control (if included in the
delivery) have to be sent to an author­
ised Ottobock Service Centre.
Conduct the self-test again by con­
necting/disconnecting
charger.
If the beep signal sounds again and
the maintenance date has not yet
been reached or exceeded, the pros­
thesis should be inspected by an
authorised Ottobock Service Centre.
The product can be used without
restrictions. However, vibration sig­
nals may not be generated.
Conduct the self-test again by con­
necting/disconnecting
charger.
If the beep/vibration signal is emitted
again, the product must be inspected
by an authorised Ottobock Service
Centre.
Connect the AXON tube adapter and then
restart the knee joint by connecting/dis­
connecting the battery charger
3
4
51%-70%
71%-90%
Genium X3 3B5-2/3B5-2=ST
been
reached
or
the
battery
the
battery
5