Mymodes; Running Function As Configured Mymode; Switching Mymodes With The Cockpit App - Otto Bock Genium X3 3B5-2 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-2:
Table of Contents

Advertisement

MyModes

11 MyModes
In addition to basic mode (mode 1), up to 5 MyModes can be activated and configured with the adjustment soft­
ware. They can be called up by the patient using the Cockpit app or remote control (optional accessory). Only the
first 3 MyModes can be selected using movement patterns. Switching by using motion patterns has to be activated
in the adjustment software.
For information on switching MyModes with the remote control (optional accessory), see the remote control
instructions for use.
These modes are intended for specific motion patterns or postures (e.g. inline skating, running (jogging) ...).
Default settings for these motion patterns and postures can be called up and individually adapted using the adjust­
ment software.
Settings can also be adjusted by the patient using the cockpit app or remote control (see Page 36).

11.1 Running function as configured MyMode

For running over longer periods of time, the "Running" MyMode which can be activated
using the cockpit app, a movement pattern or the remote control (special accessory) can be
configured using the adjustment software.
In this mode, every step will be performed as a running step with larger swing phase angle
and no preflexion at heel strike (PreFlex) (see Page 39).
INFORMATION
The running function will work with specialised running feet such as the 1E90 Sprinter and 1E95 Challenger as
well as with prosthetic feet with axial compression such as the 1C61 Triton Vertical Shock. For details on
assembly and alignment please refer to the instructions for use of the foot.
Feet without axial compression are generally not suited for running.

11.2 Switching MyModes with the cockpit app

INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the Cockpit app.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down or by connecting/discon­
necting the Bluetooth battery charger. Bluetooth is then turned on for approx. 2 minutes. During this time, the app
must be started and used to establish a connection. If required, Bluetooth on the prosthesis can be switched on
permanently afterwards (see Page 37).
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app or using the remote control (see Page 34) or if mute
mode (silent mode) is activated, there are no beep signals.
Once a connection to a prosthesis has been established, the cockpit app can be used to switch between the
MyModes.
40
Genium X3 3B5-2/3B5-2=ST

Advertisement

Table of Contents
loading

This manual is also suitable for:

Genium x3 3b5-2-st

Table of Contents