Example 1:
! Set clock frequency of /mod1 to 1MHz
! Set clock frequency of /mod2 to 5MHz
! Set clock frequency of /mod3 to 10MHz
! Set clock frequency of /mod4 to 20MHz
clockfreq device "/mod1", freq "1MHz"
clockfreq device "/mod2", freq "5MHz"
clockfreq device "/mod3", freq "10MHz"
clockfreq device "/mod4", freq" 20MHz"
Example 2:
! Use default values to set clock freq. of
! /mod1 to 1MHz
clockfreq()
Example 3:
! Set clock frequency of "/mod1" to "10MHz"
clockfreq ("/mod1","10MHz")
Remarks:
The clockfreq command sets the internal clock frequency
that is used to drive out vectors. The frequency changes
immediately upon command execution. Note that there is
no synchronization between the internal clocks of each
I/O module. Changing the internal clock frequency dur-
ing vector driving is not recommended. This frequency
is also available on the INT CLK output pin of the mod-
ule as a free-running signal and, unlike the vector output
clock, is not affected by external START, STOP, or
ENABLE conditions.
Related Commands:
syncoutput, vectordrive
clockfreq-2