Press the F3 (RENUM) key to renumber the lines while editing.
To fill a data field line when there is more than one group on the
groups line, you must enter the data with a space separating the
groups. For example, the following vector file contains two
groups that are separated by a space:
GROUPS [40-9] [8-1]
DISPLAY BIN
!TIME DATA
1 01000000000100000000011111111101 1111XXXX
Since both of the data fields in the previous example are binary,
the display line did not need to be changed. However, you could
have changed it to read:
DISPLAY BIN, BIN
and the data fields would remain the same.
If you did not enter a space on data field line in the previous ex-
ample, all of the numbers in the second group would have been
automatically tri-stated. For example, if you entered:
1111
the first field would have been right justified and filled in with
zeroes and the second field would have been tri-stated:
1 00000000000000000000000000001111 XXXXXXXX
If the display line is changed so that the second data field is a
hexadecimal number, the data that you enter on the data field
line is also expected to be in hexadecimal notation (0 through F).
For example, if the first field in the previous example was
changed to hexadecimal, and you enter the following data:
401007FD 1111XXXX