ProMinent DULCO flex Control Operating Instructions Manual page 37

Peristaltic metering pump
Table of Contents

Advertisement

Identity code 4
4
1
3
2
Fig. 23: Assignment on pump
9.2.6.4
"Current output plus relay" output (identity code C)
P_SI_0044
Fig. 24: Assignment on cable
To pin
VDE cable
1
yellow
4
green
3
white
2
brown
P_G_0073_SW
A relay combined with a current output can be ordered as an option. The
relay either switches as a fault indicating relay in the event of a fault on the
pump and with "Liquid level low 1st stage" warning messages and "Liquid
level low 2nd stage" fault messages or is used as a pacing relay.
The behaviour is factory-programmed. If another switching function is
desired, the pump can be reprogrammed in the
The variable to be signalled for the current output can be selected in the
'ANALOG OUTPUT' menu.
The current output plus relay can be retrofitted and operates once it is
plugged into the board.
Electrical interface
for current output
Data
Open circuit voltage:
Current range:
Ripple, max.:
Load, max.:
for semiconductor switch ("relay"):
Data
Max. residual voltage at I
Maximum current
Maximum voltage
Pacing pulse duration, approx.
Installation, electrical
Contact
N/O (normally open)
C (common)
N/O (normally open)
C (common)
'Relay' menu.
= 1 µA
off max
Relay
other relay
other relay
Pacing relay
Pacing relay
Value Unit
8 V
4 ... 20 mA
80 μA ss
250 Ω
Value Unit
0.4 V
100 mA
24 VDC
100 ms
37