Spectrum Analyzer (SA Mode)

SA measurements require NO calibration.
Check for a Compressed Measurement
1. Using a marker at the signal peak, make note of the
signal power level.
2. Increase the RF Attenuation level by 5 dB.
If the signal level does NOT change, then NO
-
compression exists. This indicates that the signal is
in the linear region of the receiver.
If the signal level DOES increase with more
-
attenuation, then compression DID exist. Set the
RF Attenuation value at the setting when further
increases no longer result in an increase in the
displayed power level.
Set RF Attenuation
1. Press Preset then
2. Press Mode then
3. Press Scale/Amptd then
Auto
RF Attenuation is set by adjusting the Ref
-
Level.
Man
Change the RF Attenuation level manually.
-
Display up to four types of traces
SA settings are applied to all traces.
1. Press Trace 6 then
a trace number to display.
State
2. Then
Preset (Factory)
SA
Trace 1,2,3,4
to select a trace type.
RF Atten
repeatedly to select
10