Instructions for use ELECTROmatic M/C and PM/PC
9 Maintenance | 9.1 Changing the filter - water inlet

9 Maintenance

KaVo recommends that only original KaVo parts
pairs since their safety, operation and specific suitability have been tested in ex-
tensive tests.
Note
KaVo grants no guarantee of function and KaVo accepts no liability unless
original replacement parts and operating materials are used exclusively.
Note
The device must not be serviced or repaired during treatment/use.
The servicing work described in the following can be performed by the opera-
tor/user.

9.1 Changing the filter - water inlet

Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶ Check and/or change the filter.
▶ Check the spray water channels and clean the spray nozzles with the nozzle
pin according to need.
KaVo recommends checking the filter every 3 months initially and to change the
filter insert according to need. The scope of delivery includes 2 filter inserts. Af-
ter this time, the filter should be regularly checked and replaced according to
need with the interval being appropriate for the conditions at your practice and
the degree of soiling.
▶ Undo the union nut proceeding in the direction of the arrow.
▶ Replace the filter insert (Mat. no. 1.007.9736) ① if there is any visible
soiling.