n Magazine Shutter Open
The film removing motor (MB11) rotates in the normal direction (CW) at high speed
by the removal permission signal and the suction cup arm starts moving. The removal
drive cam operates by the drive of MB11, and by moving the magazine shutter open/
close mechanism. The magazine shutter opens.
The open state of the magazine shutter is maintained by the gear cam and stopper.
016-201-01E
03.24.2006 FM4903
n Film Surface Detection
The suction cup arm passes the upper deadpoint by moving along the
groove of the removal section frame. When this passage of the upper
deadpoint is detected by the film upper deadpoint detection sensor
(SB13), MB11 switches to medium speed rotation.
When the removal conveyance roller (upper) reaches the film surface and the film
surface detection sensor (SB12) turns ON, MB11 switches to low speed rotation.
DRYPIX 2000 Service Manual
MD-24