Head Check; Fan Operation; Head Voltage.; 16Vh Ope - FujiFilm DRYPIX 2000 Service Manual

Medical dry imager
Table of Contents

Advertisement

[8]
Head Check>
[8-1]
Temperature
<FUNCTION/PURPOSE>
Command for monitoring the temperature of thermistors (THH1 to THH6).
<REMARKS>
The 100 times value of the temperature is displayed. (Unit: °C)
<OPERATIONS>
1.
Press the <UP> key/<DOWN> key, and check the measured
temperature of each thermistor.
[8-2]
Fan Operation>
<FUNCTION/PURPOSE>
Command for tuning ON/OFF the head cooling fans (FANP1, FANP2).
<OPERATIONS>
1.
Select the operations of the cooling fan.
• [ON>]:
Fan rotates.
• [OFF]:
Fan stops
• [Normal Ope.>]: Fan operates normally.
<REMARKS>
Setting to [Normal Ope.>] switches the heater to normal operations as well.
2.
If [ON>] or [Normal Ope.>] is selected, select the drive voltage.
[14V]/[18V]/[21V]/[24V]
016-201-02E
08.31.2006 FM4958
[8-3]

Head Voltage.

<FUNCTION/PURPOSE>
Command for displaying the head power voltage (+16VH) described in the individual
equipment data file.
<REMARKS>
• The second place after the decimal point is rounded up and displayed.
• The displayed voltage value varies depending on the thermal head. As a guideline,
a value in the neighborhood of 14.0 V is normal.
[8-4]
16VH Ope.>
<FUNCTION/PURPOSE>
Command for turning ON/OFF the head power (+16VH) when adjusting the head
power voltage.
<OPERATIONS>
1.
Select [ON].
→ 1 6VH is output.
2.
Adjust the head power voltage.
3.
After adjusting, select [OFF].
→ O utput of 16VH stops.
DRYPIX 2000 Service Manual
MU-44
MU-44

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Drypix lite

Table of Contents