[7-7]
Edge Sen.Mo.>
<FUNCTION/PURPOSE>
Command for monitoring the operations of the film edge sensor.
<OPERATIONS>
The procedure "n With Printing" below should be used mainly for design evaluation
purposes. In the field, this procedure should be used to log for analysis when the
film edge sensor is not working correctly. Use this function only if requested and as
instructed by the Service headquarters.
n With Printing
1.
Select the magazine slot if the sheet-feeder unit is installed.
[Slot1]/[Slot2]
→ Grid & Flat pattern will be printed.
n Checking the Operations of the Film Edge Sensor Alone
1.
[With no print] →
→ The counter value of the sensor will be displayed immediately.
2.
Confirm that the film edge sensor is active by the display of the
counter value.
<REMARKS>
The displayed counter value indicates the lateral shift of recording position in
pixels.
016-201-03E
10.18.2006 FM5036
[7-8]
Edge Sen.LED>
<FUNCTION/PURPOSE>
Command for turning ON/OFF the edge sensor LED. However, this command is not
used in most cases because the edge sensor LED is internally installed and you
cannot check its ON/OFF status in the normal assembled condition.
<OPERATIONS>
1.
Select [ON].
→ L ED of the film edge sensor lights up.
2.
Select [OFF].
→ T he LED goes off.
DRYPIX 2000 Service Manual
MU-42
MU-42