Densitometer - FujiFilm DRYPIX 2000 Service Manual

Medical dry imager
Table of Contents

Advertisement

[4-5]

Densitometer

[4-5-1] 24-Steps>
<FUNCTION/PURPOSE>
Command for printing one sheet of the 24-steps pattern required for calibration of the
built-in densitometer, and measuring the density values.
The measurement results are recorded internal for [4-5-2.Set Colle. T.>].
<REMARKS>
Correction calculation is not performed, only measurement.
<OPERATIONS>
1.
Select the magazine slot if the sheet-feeder unit is installed.
[Slot1]/[Slot2]
→ O ne sheet of the 24-steps pattern is printed.
When printing completes normally, "Completed." is displayed.
016-201-01E
06.30.2006 FM4924
[4-5-2] Set Colle. T.>
<FUNCTION/PURPOSE>
Command for creating calibration tables for calibrating measured values to match the
measured values of the built-in densitometer and external reference densitometer.
<OPERATIONS>
1.
Measure the density of each step of the 24-steps pattern printed by
[4-5-1.24-Setps>] using the external reference densiometer.
2.
Enter the density vallues (0 to 399) of each step of the pattern
measured at step 1.
→ A fter entering the density of all 24 points, pressing the <ENTER> key
executes the 24-steps density calculation.
When calibration tables are created normally, "Completed." is displayed.
<IMPLEMENTATION OF SETTINGS>
The settings made will be effective after they are executed.
DRYPIX 2000 Service Manual
MU-31
MU-31