FujiFilm DRYPIX 2000 Service Manual page 635

Medical dry imager
Table of Contents

Advertisement

[PC1-3] Setting3
[PC1-3-1] Set SMPTE
<FUNCTION>
Command for setting parameters for printing SMPTE test patterns using U-Utility.
<OPERATIONS>
1.
Select the setting number.
[1]/[2]/[3]/[4]/[5]
2.
Select [AE Title].
3.
Select [Image Matrix].
[1]/[2]/[4]/[6]/[8]/[12]
4.
Select [LUT Number].
[LUT1]/[LUT2]/[LUT3]/[LUT4]/[LUT5]/[LUT6]/[LUT7]/[LUT8]
5.
Select [Interpolation type].
[SHARP]/[MEDIUM]/[SMOOTH]
6.
Select [Interpolation algorithm method].
[CUBIC]/[NONE]
[PC1-3-2] Set Date
<FUNCTION>
Command for setting the system date and time.
<OPERATIONS>
1.
Enter the system date and time.
2.
[Execute]
016-201-06E
01.27.2012 FM5920
[PC1-3-3] Ele Save Mode
<FUNCTION>
Command for setting the timer value when this equipment shifts into the sleep mode.
<OPERATIONS>
1.
Select the sleep mode transition time.
2.
[Execute]
3.
Reboot the equipment.
[PC1-3-4] Set Alarm
<FUNCTION>
Command for setting the alarm.
<OPERATIONS>
1.
Select the alarm.
• [OFF]
: No alarm is sounded.
• [ON]
: Alarm is sounded.
• [PULSE]
: Alarm stops after about three seconds.
2.
Select the alarm tone.
[LOW]/[MEDIUM]/[HIGH]
3.
[Execute]
DRYPIX Lite Service Manual
MU-46
MU-46